This is a technical guide from Gentec-EO for setting up and using their THz-B-DA and THz-I-BNC Analog Radiometer Probes. These THz detectors are pyroelectric, high-sensitivity "thermal" detectors designed for use with an Oscilloscope or a Lock-In Amplifier (LIA). The voltage output is maximized at a 5 Hz chopping frequency.

Key Setup Considerations

  • Avoid heat sources (like power supplies or human bodies) as the probes are very sensitive to infrared (IR) radiation.

  • The detectors are also sensitive to acoustic waves and mechanical vibration ("piezoelectric" properties), so avoid air conditioning vents and mechanical pumps.

  • An optical chopper running at 5 Hz is required for operation.

Powering the Probes

  • THZ-I-BNC probes include batteries and an AC voltage supply.

  • THZ-B-DA probes require the external battery and AC supply T-Rad-Analog.

  • Both models can work off batteries or an AC wall wart.

Step-by-Step Instructions (Using an Oscilloscope)

  1. Mount and Align the probe to the optical axis of the THz source.

  2. Power Up the probe using the rocker switch; a green power light should illuminate.

  3. Connect the BNC analog output to the oscilloscope.

  4. Wait for Thermal Equilibrium: When first powered up, the output may be a large negative voltage (e.g., $-10$ V), but it will stabilize to a final value of +200 mV after about 5 minutes.

  5. Expose the Probe to your source by unblocking the beam.

  6. Measure Power: Adjust the scope to get a good sawtooth voltage output (due to the chopper).

    • Measure the peak voltage and subtract the base voltage (this accounts for thermal drift).

    • Divide the result by the probe's voltage responsivity ($R_v$) in V/W to determine the power in Watts.

    • Ensure the detector is not being overfilled.

The document also provides diagrams for setting up the THz detector with an oscilloscope and with a Lock-In Amplifier, noting that the LIA setup is suitable for lower power levels and requires using the chopper's "sync" output as the reference.

access the Article!

Get unlimited access to:

Trend and Thought Leadership Articles
Case Studies & White Papers
Extensive Product Database
Members-Only Premium Content
Welcome Back! Please Log In to Continue. X

Enter your credentials below to log in. Not yet a member of Photonics Online? Subscribe today.

Subscribe to Photonics Online X

Please enter your email address and create a password to access the full content, Or log in to your account to continue.

or

Subscribe to Photonics Online

Gentec Electro-Optics, Inc.